Research article
07 Jun 2017
Methods to homogenize electrochemical concentration cell (ECC) ozonesonde measurements across changes in sensing solution concentration or ozonesonde manufacturer
Terry Deshler1, Rene Stübi2, Francis J. Schmidlin3, Jennifer L. Mercer1,a, Herman G. J. Smit4, Bryan J. Johnson5, Rigel Kivi6, and Bruno Nardi1,7 1Department of Atmospheric Science, University of Wyoming, Laramie, Wyoming, USA
2Meteo Swiss, Payerne, Switzerland
3NASA Goddard Space Flight Center, Greenbelt, Maryland, USA
4Forschungszentrum Jülich GmbH, Institute of Energy and Climate Research: Troposphere, Jülich, Germany
5National Oceanic and Atmospheric Administration, Boulder, Colorado, USA
6Finnish Meteorological Institute, Arctic Research, Sodankylä, Finland
7Nardi Scientific, LLC, Denver, Colorado, USA
anow at: National Science Foundation, Geosciences Directorate, Division of Polar Programs, Arlington, Virginia, USA
Abstract. Ozone plays a significant role in the chemical and radiative state of the atmosphere. For this reason there are many instruments used to measure ozone from the ground, from space, and from balloons. Balloon-borne electrochemical cell ozonesondes provide some of the best measurements of the ozone profile up to the mid-stratosphere, providing high vertical resolution, high precision, and a wide geographic distribution. From the mid-1990s to the late 2000s the consistency of long-term records from balloon-borne ozonesondes has been compromised by differences in manufacturers, Science Pump (SP) and ENSCI (EN), and differences in recommended sensor solution concentrations, 1.0 % potassium iodide (KI) and the one-half dilution: 0.5 %. To investigate these differences, a number of organizations have independently undertaken comparisons of the various ozonesonde types and solution concentrations, resulting in 197 ozonesonde comparison profiles. The goal of this study is to derive transfer functions to allow measurements outside of standard recommendations, for sensor composition and ozonesonde type, to be converted to a standard measurement and thus homogenize the data to the expected accuracy of 5 % (10 %) in the stratosphere (troposphere). Subsets of these data have been analyzed previously and intermediate transfer functions derived. Here all the comparison data are analyzed to compare (1) differences in sensor solution composition for a single ozonesonde type, (2) differences in ozonesonde type for a single sensor solution composition, and (3) the World Meteorological Organization's (WMO) and manufacturers' recommendations of 1.0 % KI solution for Science Pump and 0.5 % KI for ENSCI. From the recommendations it is clear that ENSCI ozonesondes and 1.0 % KI solution result in higher amounts of ozone sensed. The results indicate that differences in solution composition and in ozonesonde type display little pressure dependence at pressures  ≥  30 hPa, and thus the transfer function can be characterized as a simple ratio of the less sensitive to the more sensitive method. This ratio is 0.96 for both solution concentration and ozonesonde type. The ratios differ at pressures < 30 hPa such that OZ0. 5%/OZ1. 0 % =  0. 90 + 0. 041 ⋅ log10(p) and OZSciencePump/OZENSCI =  0. 764 + 0. 133 ⋅ log10(p) for p in units of hPa. For the manufacturer-recommended solution concentrations the dispersion of the ratio (SP-1.0 / EN-0.5 %), while significant, is generally within 3 % and centered near 1.0, such that no changes are recommended. For stations which have used multiple ozonesonde types with solution concentrations different from the WMO's and manufacturer's recommendations, this work suggests that a reasonably homogeneous data set can be created if the quantitative relationships specified above are applied to the non-standard measurements. This result is illustrated here in an application to the Nairobi data set.

Citation: Deshler, T., Stübi, R., Schmidlin, F. J., Mercer, J. L., Smit, H. G. J., Johnson, B. J., Kivi, R., and Nardi, B.: Methods to homogenize electrochemical concentration cell (ECC) ozonesonde measurements across changes in sensing solution concentration or ozonesonde manufacturer, Atmos. Meas. Tech., 10, 2021-2043, https://doi.org/10.5194/amt-10-2021-2017, 2017.
Publications Copernicus
Download
Short summary
Ozonesondes, small balloon-borne instruments to measure ozone profiles, are used once and lost. Quality control is thus essential. From the mid-1990s to late 2000s differences in manufacturers' (Science Pump and ENSCI) recommended sensor solution concentrations, 1.0 % and 0.5 % potassium iodide, led to some confusion. This paper uses comparison measurements to derive transfer functions to homogenize the measurements made with non-standard combinations of instrument and sensor solution.
Ozonesondes, small balloon-borne instruments to measure ozone profiles, are used once and lost....