Articles | Volume 10, issue 8
https://doi.org/10.5194/amt-10-3073-2017
https://doi.org/10.5194/amt-10-3073-2017
Research article
 | 
24 Aug 2017
Research article |  | 24 Aug 2017

Effects of variation in background mixing ratios of N2, O2, and Ar on the measurement of δ18O–H2O and δ2H–H2O values by cavity ring-down spectroscopy

Jennifer E. Johnson and Chris W. Rella

Viewed

Total article views: 2,473 (including HTML, PDF, and XML)
HTML PDF XML Total BibTeX EndNote
1,628 747 98 2,473 80 89
  • HTML: 1,628
  • PDF: 747
  • XML: 98
  • Total: 2,473
  • BibTeX: 80
  • EndNote: 89
Views and downloads (calculated since 24 Apr 2017)
Cumulative views and downloads (calculated since 24 Apr 2017)

Viewed (geographical distribution)

Total article views: 2,473 (including HTML, PDF, and XML) Thereof 2,440 with geography defined and 33 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 

Cited

Latest update: 18 Apr 2024
Download
Short summary
Cavity ring-down spectrometers (CRDS) can make accurate measurements of trace gases when the inert gas background is constant but are prone to errors when the background is variable. Here, we report that in one CRDS analyzer (L2120-i, Picarro), these errors result from broadening, narrowing, and shifting of the target absorption lines as well as neighboring lines. New spectral acquisition and analysis strategies are needed to make CRDS measurements robust to variable backgrounds.