Journal metrics

Journal metrics

  • IF value: 3.248 IF 3.248
  • IF 5-year value: 3.650 IF 5-year 3.650
  • CiteScore value: 3.37 CiteScore 3.37
  • SNIP value: 1.253 SNIP 1.253
  • SJR value: 1.869 SJR 1.869
  • IPP value: 3.29 IPP 3.29
  • h5-index value: 47 h5-index 47
  • Scimago H index value: 60 Scimago H index 60
Volume 10, issue 12 | Copyright
Atmos. Meas. Tech., 10, 5063-5073, 2017
https://doi.org/10.5194/amt-10-5063-2017
© Author(s) 2017. This work is distributed under
the Creative Commons Attribution 3.0 License.

Research article 22 Dec 2017

Research article | 22 Dec 2017

Improved methods for signal processing in measurements of mercury by Tekran® 2537A and 2537B instruments

Jesse L. Ambrose
Data sets

Data Archive for: Improved methods for signal processing in measurements of mercury by Tekran® 2537A and 2537B instruments J. Ambrose https://doi.org/10.17605/OSF.IO/KGFYP

Publications Copernicus
Download
Short summary
Scientific understanding of environmental Hg cycling is limited by analytical uncertainties. To better characterize analytical uncertainty associated with Hg measurements made with the Tekran® 2537 instrument, I developed new software-based methods for offline processing of the raw instrumental data. I demonstrate significant uncertainty associated with the Tekran® method. By comparison, my methods improve measurement accuracy and the Hg detection limit by as much as 95 % and 88 %, respectively.
Scientific understanding of environmental Hg cycling is limited by analytical uncertainties. To...
Citation
Share