Preflight calibration of the Chinese Environmental Trace Gases Monitoring Instrument (EMI)

Min Jie Zhao, Fu Qi Si, Hai Jin Zhou, Shi Mei Wang, Yu Jiang, and Wen Qing Liu

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
AR by MinJie Zhao on behalf of the Authors (05 Sep 2018)  Author's response    Manuscript
ED: Publish as is (07 Sep 2018) by Hendrik Fuchs
Download
Short summary
EMI is a nadir-viewing wide-field imaging spectrometer aiming to quantify the global distribution of tropospheric and stratospheric trace gases. The spectral range is 240–710 nm, with the spectral resolution 0.3–0.5 nm. The preflight accuracy of wavelength calibration is less than 0.05 nm; the accuracy of radiance calibration is less than 5 %. The obtained calibration key data are used for the L1b processor. In-orbit wavelength/radiometric calibration is used to monitor performance changes.