Articles | Volume 12, issue 5
https://doi.org/10.5194/amt-12-2665-2019
https://doi.org/10.5194/amt-12-2665-2019
Research article
 | 
09 May 2019
Research article |  | 09 May 2019

Characterization of a commercial lower-cost medium-precision non-dispersive infrared sensor for atmospheric CO2 monitoring in urban areas

Emmanuel Arzoumanian, Felix R. Vogel, Ana Bastos, Bakhram Gaynullin, Olivier Laurent, Michel Ramonet, and Philippe Ciais

Viewed

Total article views: 3,435 (including HTML, PDF, and XML)
HTML PDF XML Total Supplement BibTeX EndNote
2,045 1,320 70 3,435 232 88 79
  • HTML: 2,045
  • PDF: 1,320
  • XML: 70
  • Total: 3,435
  • Supplement: 232
  • BibTeX: 88
  • EndNote: 79
Views and downloads (calculated since 04 Oct 2018)
Cumulative views and downloads (calculated since 04 Oct 2018)

Viewed (geographical distribution)

Total article views: 3,435 (including HTML, PDF, and XML) Thereof 3,166 with geography defined and 269 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 

Cited

Latest update: 19 Apr 2024
Download
Short summary
We tested commercial lower-cost CO2 sensors in laboratory and field studies to see if they can measure atmospheric CO2 mole fractions with less than 1 ppm bias (with monthly calibration), to allow continuous urban CO2 monitoring. We find that the sensors' CO2 readings are influenced by temperature, atmospheric pressure and water vapour content, but this can be corrected for by adding sensors (T, p, RH) and carefully calibrating each sensor against a high-precision instrument.