Articles | Volume 12, issue 7
https://doi.org/10.5194/amt-12-3907-2019
https://doi.org/10.5194/amt-12-3907-2019
Research article
 | 
16 Jul 2019
Research article |  | 16 Jul 2019

Laboratory and field evaluation of the Aerosol Dynamics Inc. concentrator (ADIc) for aerosol mass spectrometry

Sanna Saarikoski, Leah R. Williams, Steven R. Spielman, Gregory S. Lewis, Arantzazu Eiguren-Fernandez, Minna Aurela, Susanne V. Hering, Kimmo Teinilä, Philip Croteau, John T. Jayne, Thorsten Hohaus, Douglas R. Worsnop, and Hilkka Timonen

Viewed

Total article views: 2,150 (including HTML, PDF, and XML)
HTML PDF XML Total Supplement BibTeX EndNote
1,401 692 57 2,150 298 48 59
  • HTML: 1,401
  • PDF: 692
  • XML: 57
  • Total: 2,150
  • Supplement: 298
  • BibTeX: 48
  • EndNote: 59
Views and downloads (calculated since 15 Mar 2019)
Cumulative views and downloads (calculated since 15 Mar 2019)

Viewed (geographical distribution)

Total article views: 2,150 (including HTML, PDF, and XML) Thereof 1,875 with geography defined and 275 with unknown origin.
Country # Views %
  • 1
1
 
 
 
 

Cited

Latest update: 24 Apr 2024
Download
Short summary
An air-to-air ultrafine particle concentrator (Aerosol Dynamics Inc. concentrator; ADIc) has been tailored for the low (~ 0.08 L min−1) inlet flow of aerosol mass spectrometers, and it provides a factor of 8–21 enrichment in the concentration of particles. The ADIc was evaluated in laboratory and field measurements. The results showed that the concentration factor depends primarily on the ratio between the sample flow and the output flow and is independent of particle size above about 10 nm.