Articles | Volume 12, issue 10
https://doi.org/10.5194/amt-12-5335-2019
https://doi.org/10.5194/amt-12-5335-2019
Research article
 | 
08 Oct 2019
Research article |  | 08 Oct 2019

Microelectromechanical-system-based condensation particle counter for real-time monitoring of airborne ultrafine particles

Seong-Jae Yoo, Hong-Beom Kwon, Ui-Seon Hong, Dong-Hyun Kang, Sang-Myun Lee, Jangseop Han, Jungho Hwang, and Yong-Jun Kim
Abstract

We present a portable, inexpensive, and accurate microelectromechanical-system-based (MEMS-based) condensation particle counter (CPC) for sensitive and precise monitoring of airborne ultrafine particles (UFPs) at a point of interest. A MEMS-based CPC consists of two main parts: a MEMS-based condensation chip that grows UFPs to micro-sized droplets and a miniature optical particle counter (OPC) that counts single grown droplets with the light scattering method. A conventional conductive cooling-type CPC is miniaturized through MEMS technology and three-dimensional (3-D) printing techniques; the essential elements for growing droplets are integrated on a single glass slide. Our system is much more compact (75 mm × 130 mm × 50 mm), lightweight (205 g), and power-efficient (2.7 W) than commercial CPCs. In quantitative experiments, the results indicated that our system could detect UFPs with a diameter of 12.9 nm by growing them to micro-sized (3.1 µm) droplets. Our system measured the UFP number concentration with high accuracy (mean difference within 4.1 %), and the number concentration range for which our system can count single particles is 7.99–6850 cm−3. Thus, our system has the potential to be used for UFP monitoring in various environments (e.g., as an air filtration system, in high-precision industries utilizing clean rooms, and in indoor and outdoor atmospheres).

Dates
1 Introduction

Monitoring of airborne ultrafine particles (UFPs, which are smaller than 100 nm) is needed in various fields for human health and yield enhancement in industrial fields (Donaldson et al., 1998; Donovan et al., 1985; Hristozov and Malsch, 2009; Li et al., 2016; Liu et al., 2015). While UFPs have a variety of anthropogenic and natural sources (e.g., soot agglomerates and secondary particles from hazardous gaseous precursors), in urban areas they are largely generated from vehicle exhaust (Kim et al., 2011; Kittelson, 1998; Shi et al., 1999). Because of dramatic developments in nanotechnology, engineered UFPs for commercial and research purposes have been produced on a large scale. These incidentally and intentionally generated UFPs are more harmful to human health than their larger counterparts. These UFPs have a higher chance of being deposited in the lower respiratory system and are more toxic owing to their larger surface-to-volume ratios, which causes oxidative stress, pulmonary inflammation, and tumor development (Hesterberg et al., 2012; Hext, 1994; Li et al., 2003; Renwick et al., 2004). Thus, on-site monitoring is needed to assess and minimize UFP exposure. High-precision industries with clean rooms also require UFP monitoring to trace their sources, minimize the contamination, and thereby increase the production yield. For instance, in the semiconductor industry, where the minimum feature size of the semiconductor devices approaches 7 nm, particles with diameters of a few nanometers are critical (Neisser and Wurm, 2015). Although an air-purification system equipped with an ultra-low particulate filter eliminates the contaminants in the air entering the clean room, it cannot control the UFPs internally generated during the manufacturing processes (e.g., chemical vapor deposition (CVD), metallization, and wet etching) (Choi et al., 2015; Manodori and Benedetti, 2009). If they are deposited on the electrodes of a chip, they cause interruption of the current flow, making the whole chip unusable and thereby reducing the yield of the semiconductor production (Libman et al., 2015). In this regard, ISO 14644-12 has been recently developed to provide guidance on how to monitor UFPs in clean rooms. To monitor the concentration field of UFPs in these environments where the spatial and temporal variations of UFP concentrations are enormous, portable and low-cost sensors are required to establish simultaneous monitoring at multiple points or dense monitoring networks.

Condensation particle counters (CPCs) are the most widely used UFP detection instruments and are based on the heterogeneous particle condensation technique (Stolzenburg and McMurry, 1991). They grow UFPs to micro-sized droplets through condensation and count them by optical means. Compared to the electrical method (measuring the number concentration of UFPs by electrically charging them and sensing their current), CPCs provide extremely sensitive and precise counting because they are capable of counting individual particles (Kangasluoma et al., 2014, 2017; McMurry, 2000). Moreover, if a differential mobility analyzer (DMA) is used as a particle size selector, CPCs can offer higher particle size resolution than any other particle sizing instruments (Sioutas, 1999; Stolzenburg et al., 2017). However, commercially available CPCs are bulky and expensive; thus, they are impractical for on-site monitoring where the UFP concentration changes continuously. Although portable CPCs (e.g., model 3007, TSI Inc., USA) are currently on the market, they are still large in size (292 mm × 140 mm × 140 mm) and expensive ( USD 10 000). Therefore, despite their advantages, CPCs are difficult to utilize actively for on-site monitoring applications.

In the past few years, several studies have used microelectromechanical-system (MEMS) technology for UFP analysis (Hajjam et al., 2010; Kang et al., 2012; Kim et al., 2015, 2018; Wasisto et al., 2013; Zhang et al., 2016). Because this technology is capable of batch and micro-size fabrications through the semiconductor manufacturing process, such chips provide cost efficiency, compactness, and enhanced portability. Most MEMS-based sensors developed for monitoring UFPs were based on the electrical technique. However, these sensors exhibited low sensitivity because their small size means that they have to operate at a small volumetric flow rate (below 1 L min−1), and UFPs carry only a small number of elementary charges of their limited surface area.

In this study, we developed a high-performance MEMS-based CPC that is portable, inexpensive, and power-efficient. Our system comprises a MEMS-based condensation chip and miniature optical particle counter (OPC). UFPs are grown to micrometer-sized droplets on the chip, and the grown droplets are detected by the miniature OPC. New fabrication techniques including MEMS and 3-D printing technologies are applied to CPCs; particle enlargement (i.e., the fundamental process of the CPC) can be realized on a chip-scale system because the essential elements for growing droplets (i.e., the channels, micropillar-type wick, heaters, and temperature sensors) are integrated on a single glass slide. Accordingly, our system is far more compact and cost-efficient than traditional CPCs, even when considering the portable versions.

In addition to its compactness, our system also provides high degrees of accuracy and precision. A quantitative characterization using Ag particles proves that our system is capable of growing UFPs to micrometer-sized droplets, counting them one by one, and thereby measuring UFP number concentration with a high accuracy, which is comparable to a commercial OPC. These results show that our system can potentially be used as a portable, low-cost, and high-precision UFP sensor for various fields (e.g., assessing UFP exposure, monitoring workplaces, and tracing particle sources in high-precision industries with clean rooms). Moreover, when combined with the recently developed miniature DMA, our system should also be able to perform high-resolution on-site monitoring of the UFP size distribution (Liu and Chen, 2016; Qi and Kulkarni, 2016).

2 Description of the MEMS-based CPC

Figure 1 shows the operating principle of the proposed MEMS-based CPC, which consists of a reservoir, saturator, condenser, and miniature OPC. To generate supersaturated vapor and hence to grow UFPs to micro-sized droplets, our system utilizes a conductive cooling method, with butanol as the working fluid. The saturator generates saturated vapor by heating the wetted wall with the working fluid. This saturates the UFP-laden sample, which is introduced into the saturator at a flow rate of 0.15 L min−1, with working fluid vapor. The saturated sample then enters a condenser, whose temperature (10 C) is lower than that of the saturator (40 C). In the condenser, the hot saturated vapor present in the sample is cooled to reach the supersaturated state. UFPs in the supersaturated vapor act as condensation nuclei and grow to droplets.

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f01

Figure 1Schematic illustration of the MEMS-based CPC. Our system consists of four parts: the reservoir, saturator, condenser, and miniature OPC. The reservoir supplies the working fluid to the saturator via capillary action by the micropillar-type wick. The saturator heats the working fluid to generate saturated vapor. The saturated air becomes supersaturated when cooled by the condenser. UFPs grow into micro-sized droplets in the condenser. They are counted by the miniature OPC.

Download

The cornerstone of our system is its compact, cost-efficient, and portable features for the on-site monitoring of UFPs. Figure 2a shows our system with the customized circuit. By using the MEMS technology, our system can generate supersaturated vapor and grow droplets on a chip-scale system for significant decreases in size, weight, and power consumption. The circuit, whose dimensions are 90 mm × 65 mm, simultaneously reads the data from the miniature OPC, temperature sensor, and flow sensor (model FS1012-1020-NG, IDT Co., USA), and it controls the power of the heaters, cooling modules, and micro pump (model 00H220H024, Nidec Co., Japan) via a pulse-width modulation (PWM) method. In order for our system to be a stand-alone device, the feedback loops based on the proportional–integral–differential (PID) algorithm are implemented in the micro control unit of the circuit, and their gains can easily be controlled using serial communication. The heaters, temperature sensors, and wick are monolithically integrated with the glass slide (Fig. 2b). They perform crucial roles: supplying the working fluid to the saturator and condenser via capillary action and generating supersaturated vapor.

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f02

Figure 2(a) Optical image of the MEMS-based CPC. (b) Magnified image of the heaters, resistive temperature sensors, and wick on the glass slide. (c) Image from a scanning electron microscope (SEM) of the micropillar-type wick.

Download

As shown in Fig. 2c, micropillar arrays serve as a wick with a 40 µm diameter, 100 µm length, and 100 µm pitch. The dimensions of the micropillar-type wick were experimentally determined to be capable of pumping the working fluid from the reservoir and spreading it over the entire surface of the saturator to ensure that the saturator wall is always in the wetted condition. SU-8 is a negative-tone photoresist and was chosen as the structural material for the micropillar-type wick because it provides a high patterning resolution ( 1 µm) and outstanding chemical/thermal stability and guarantees the high durability of our system (Chang and Kim, 2000; Kim et al., 2003).

In order to generate supersaturated vapor with a constant saturation ratio, the temperatures of the saturator (40 C) and condenser (10 C) must be controlled to the designed values. For this purpose, the resistive heater was uniformly patterned on the inner wall of the saturator, and a miniature thermoelectric cooling module was attached to the outer wall of the condenser. Both temperatures were monitored with resistive temperature sensors having an accuracy of ±0.1C and located at the outlet of the saturator and center of the condenser, respectively. A customized circuit was used with the PWM method to adjust the power for the heaters and thermoelectric cooler module and thus to control the temperatures.

In the condenser, while the supersaturated vapor grows UFPs to droplets, the working fluid vapor may condense on the wall and clog the channel. Thus, like the saturator, the condenser also had micropillar-type wicks. On a rough surface, the actual contact angle of a working fluid droplet is lower than the contact angle on a smooth surface (Chen et al., 2013). Thus, the micropillar array increases the wettability of the working fluid, suppressing the droplet formation on the wall and draining the condensed working fluid to the reservoir. While the diameter and length were the same, the pitch of the micropillar-type wick (130 µm) was larger than that in the saturator (100 µm).

A spacer, which included the channels and inlet/outlet connectors, was fabricated with a 3-D printer during a single printing. The material for the spacer was UV-curable epoxy, which has a high printing resolution (minimum linewidth: 0.3 mm) and can endure temperatures up to 80 C (Stansbury and Idacavage, 2016). The channel of the saturator was winding to increase the residence time and thus ensure fully saturated vapor. The width, height, and length of the saturator channel were 6, 3, and 150 mm, respectively, and the corresponding residence time at the given flow rate was 1.08 s. The maximum Reynolds number in the channel at the given flow rate was only 32, which means that the sample stream in the channel was in the fully laminar regime.

The detection part of the commercial OPC (Innoair-615D, Innociple Co., Republic of Korea), which provided a time resolution of 6 s, was used as the optical detector in our system. It consists of the sensing chamber and optics (laser, cylindrical lens, elliptic mirror, photodiode, and light trap). The introduced droplets are arranged in a row in the acceleration nozzle (i.e., the outlet of the condensation chip). Then they enter the sensing chamber. The droplets then pass through the place where the condensed thin beam is irradiated. The mirror directs the light scattered from a droplet to the sensing surface of the photodiode. When the laser beam passes through the cylindrical lens, the shape of the laser beam at the focal point is not a point but a very thin surface. In addition, the acceleration nozzle at the chip outlet is only 0.8 mm in diameter and is located about 1.5 mm below the point through which the beam passes. Therefore, under the condition that there is no coincidence error (the two particles do not pass through the viewing volume at the same time), almost all the grown micro-droplets are counted in the optical detector.

3 Fabrication process

As shown in Fig. 3, the MEMS-based particle growth system consists of a top plate, bottom plate, and 3-D-printed channel. The fabrication process was identical for the two plates. The essential elements on the plates (heaters, resistive temperature sensors, and the micropillar-type wick) were fabricated through a simple photolithographic process. An electron beam evaporator was used to deposit a thin metal layer (30∕300 nm of Ti  Au) on the plates. Then, a positive-tone photoresist was spin-coated at 3000 rpm for 30 s, and it softly baked on a hot plate at 95 C for 1 min. After baking, the photoresist was exposed to UV light (wavelength: 365 nm, exposure dose: 55 mJ cm−2) and developed with a photoresist developer. Then, the wet etching process was performed to define the electrodes (Fig. 3a). To enhance the repeatability of the temperature sensors, the fabricated electrodes were heat-treated at 300 C in ambient atmosphere. The SU-8 negative-tone photoresist was used as the material for the micropillar arrays. A 100 µm layer of SU-8 (model 2100, Microchem Corp., USA) was spin-coated onto the plates at 3000 rpm for 30 s and baked at 85 C for 120 min. Then, it was exposed to UV light (exposure dose: 240 mJ cm−2) to define the micropillar array structure. Next, a post-exposure bake (PEB) was conducted through a two-step ramping process on a hot plate: 65 C for 5 min and then 95 C for 120 min. The exposed SU-8 was developed with an SU-8 developer (Fig. 3b).

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f03

Figure 3Simplified fabrication process of the MEMS-based CPC.

Download

The fabricated channel and plates were packaged with polymethyl methacrylate jigs and silicon rubber gaskets (Fig. 3c). The cooling modules comprised a thermoelectric cooler, heat sink, and mini fan. They were attached to the outer walls of the condenser (Fig. 3d). Finally, the fabricated chip was inserted into a card connector, which was linked to the control circuit board.

4 Experimental setup

Figure 4 shows the experimental setup used to characterize the overall performance of our system in terms of three aspects: (a) the clean air supply system, (b) monodisperse-particle-generating system, and (c) performance comparison system. Compressed air was used as the carrier gas. Any moisture, oil droplets, and particles in the compressed air were removed in the clean air supply system with an oil trap, diffusion dryer, and high-efficiency particulate air (HEPA) filter. The purified air was then supplied to the particle generating system at a flow rate that was accurately controlled by a mass flow controller (VIC-D200, MKP Co., Republic of Korea). Ag particles ranging in size from 3 to 140 nm were generated by an Ag particle generator (EP-NGS20, EcoPictures Co., Republic of Korea). The particles were electrically charged by a soft X-ray charger (XRC-05, HCT Co., Republic of Korea). Then they were classified according to diameter with two types of DMAs: (1) nano DMA (model 3085, TSI Co. Ltd., USA) for particles in the size range from 3 to 10 nm or (2) long DMA (model 3081A, TSI Co. Ltd., USA) for particles in the size range from 5 to 140 nm. Next, the number concentration of the monodisperse Ag particles was controlled (0–24 000 cm−3) in the dilution bridge system by adjusting the needle valve. Finally, the concentration-controlled and monodisperse Ag particles were introduced into our system and reference instrument, which was either a CPC (model 3772, TSI Inc., USA) or an aerosol electrometer (model 3068B, TSI Inc., USA).

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f04

Figure 4Schematic of the experimental setup for evaluating the performance of the MEMS-based CPC.

Download

Because of the large difference between the flow rates of the reference instrument and our system, the following procedures were carried out to verify that particles with the same concentration were introduced into the two systems. First, to minimize the particle loss induced from the turbulence at the bifurcation, a flow splitter with a very small angle of split (model 3708, TSI Inc., USA) was used. The tubes which lead to both systems were electrostatically dissipative to minimize the electrostatic particle loss, and their lengths were carefully adjusted to match the transportation times. To verify that the particles which were introduced into both systems had the same concentrations, it was confirmed that the counting efficiency was close to 100 % when particles with a size of 100 nm were introduced (it was assumed that they were activated and grew into droplets with 100 % efficiency). Then, while reducing the size of the introduced particles to 40 nm by adjusting the voltages of the DMA, it was confirmed that the counting efficiency remained constant. Through these procedures, it was verified that the concentrations of the particles delivered to the two systems were the same. The loss of our system was characterized using the counting efficiency, since it is defined as the efficiency of the system at detecting the introduced particles, thereby describing the overall transportation and activation efficiencies.

5 Results and discussion

5.1 Working-fluid transmission and evaporation

In order to characterize the wick capability of the working-fluid transmission and evaporation, a rectangular test sample was used. The sample included a heater and temperature sensor, and its length was equal to the saturator maximum length of our system (38 mm).

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f05

Figure 5(a) Schematic of the capillary rise experimental setup. (b) Selected video frames from the rise of the working fluid using the micropillar-type wick. (c) The dry-out region formation as the surface temperature increased.

Download

To identify the supply capability of the wick, a capillary rise experiment was performed (Fig. 5a). The sample was slowly lowered to the surface of the working fluid. Once the bottom of the sample touched the surface (t=0 s), the rise of the working fluid on the sample was recorded. As shown in Fig. 5b, the forefront of the working fluid rose rapidly at the beginning and then slowed down gradually. The forefront of the working fluid reached the endpoint of the saturator in 35 s and thus wetted the whole surface of the sample.

In the saturator, the working fluid transported by the wick is vaporized at elevated temperatures. When the rate of the working-fluid evaporation is higher than that of the transportation, a dry-out region is formed. This phenomenon must be minimized because the partial vapor pressure near the wall must be kept close to the saturated condition. Figure 5c shows optical images of the dry-out region formation as the surface temperature increased. The temperature of the sample surface was increased in increments of 10 C; at each temperature, the system operated for 180 s. The dry-out region clearly did not form when the surface temperature was equal to the designed saturator temperature (40 C) or even when it reached 70 C. At 80 C, the front of the working fluid started to recede, so a dry-out region formed. At 90 C, the area of the dry-out region accounted for approximately 20 % of the heated surface. These results demonstrate that the amount of the working fluid supplied from the wick was greater than the rate of evaporation in the condenser at the design temperature. Moreover, the saturator temperature can be raised to 70 C to increase the saturation ratio and hence decrease the Kelvin diameter in order to enable the system to detect smaller particles.

In order to obtain optical images of the dry-out region formation, a single glass slide with patterned electrodes and micropillar array was used. For this reason, there was no flow rate during the experiment. Although, due to the advection, the flow rate affects the area of the dry-out region to some extent, the vapor pressure at which the dry-out region started to form (164.1 mmHg at 80 C) was 8.7 times larger than the designed value of the saturator (18.9 mmHg at 40 C). Thus, although there was no flow above the butanol surface when measuring the dry-out region, the dry-out region did not occur in the saturator of the MEMS-based CPC under operating conditions.

5.2 Droplet size distribution

Figure 6 shows the size distribution of the droplets generated from the MEMS-based particle growth system. Monodisperse Ag particles in the size range from 20 to 140 nm were used as a test aerosol, and their number concentrations were fixed at around 2000 cm−3 by adjusting the valves of the dilution bridge. The sampling time for measuring each droplet distribution was 2 min, and the corresponding measurement uncertainty based on Poisson statistics was 0.13 %. All the error bars at each data point represent the standard deviations. A commercial OPC (OPC-N2, Alphasense, UK) was used for measuring the droplet size distribution. It has been reported that OPC-N2 was capable of not only measuring particles from 0.4 to 17 µm, but it was also capable of having moderate counting performance compared to the reference OPC (PAS-1.108, Grimm Technologies) (Sousan et al., 2016). The measurement errors induced from the Mie resonance were not considered in these data. The average droplet diameter (dd,avg) was 3.1 µm when particles of 20 nm in size, slightly larger than the minimum detectable size (12.9 nm), were introduced. Because the lower detectable size of the optical detector in our system was 0.3 µm, the introduced particles successfully grew into micrometer-sized droplets that were large enough to be counted by optical means. It was noted that the mean droplet size did not vary significantly above 40 nm. In addition, most of the grown droplets were smaller than 10 µm, indicating that droplets tens of micrometers in size, which could attach to the inner walls of the particle growth system or optical detector via sedimentation, were barely generated.

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f06

Figure 6Size distribution of the droplets grown from the MEMS-based particle growth system when Ag particles of specific sizes were introduced.

Download

5.3 Size-dependent particle counting efficiency

The counting efficiency (ηd) is defined as the efficiency of the system at detecting the particles and describes the overall CPC performance. It is the product of three efficiencies:

(1) η d = η trans η act η OPC ,

where ηtrans is the efficiency of a particle passing through our system, ηact is the efficiency of growing droplets at the condensation chip, and ηOPC is the efficiency of counting droplets passing through the sensing volume. Because these three efficiencies are strongly dependent on the particle size, in particular for small particles below ca. 30 nm, the counting efficiency must be characterized as a function of the particle size.

The counting efficiency was obtained from the ratio of the concentration measured with our system to the reference number concentration. The reference concentration (Nref) was obtained from the electrical current (I) measured by the aerosol electrometer (AE).

(2) I = N ref n e Q ,

where n is the number of elementary charges (+1) per particle, e is the elementary charge (1.6×10-19 C), and Q is the flow rate drawn into the AE.

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f07

Figure 7Particle counting efficiency of the MEMS-based CPC as a function of the particle size and saturator temperature. The particle size at which the particle counting efficiency is 50 % was 12.9 nm (TS=40C), 17.3 (TS=35C), and 20.4 (TS=30C), respectively.

Download

Figure 7 shows the size-dependent counting efficiency of the MEMS-based CPC. The size range of the Ag particles was controlled so that the concentration range was 1000–2000 cm−3. The sampling times for each data point were 300 s, and the measurement uncertainty based on Poisson statistics was 0.02 %. To evaluate the effect of the temperature difference, the counting efficiency was characterized when the condenser temperature (Tc) was 10 C and the saturator temperatures (Ts) were 30, 35, and 40 C. At 40 C (the design value of the saturator temperature), the same experiments were repeated three times to confirm the measurement reliability. When the saturator temperature was 40 C, it was found that our system detected 1 % of UFPs with size 5 nm, and the detection efficiency increased sharply above 9 nm. This was primarily because the activation efficiency (ηact) increased when the particle size exceeded the Kelvin diameter (2.34 nm). The transport efficiency (ηtrans) also increased, because the diffusivity of a particle decreases with increasing particle size. The counting efficiency data were curve-fitted using

(3) η d = α + ( β - α ) 1 + d p / γ δ ,

where α, β, γ, and δ are fitting constants with values of 101.96, 2.00, 12.99, and 4.70, respectively. The corresponding minimum detectable size is defined as the size at which particles are detected with 50 % efficiency. It was found to be 12.9 nm. The detection efficiency was 90 % at 20.1 nm and reached 95 % at 22.9 nm. It was close to 100 % and constant in the size range from 25 to 140 nm, indicating that the internal particle loss in this size range was negligible.

The minimum detectable size was higher than that of the commercial CPC operating at the same temperature difference. Because the saturator and condenser were close to each other and thereby have thermal interference, the condenser temperature might be higher than its originally designed value due to the heat transfer from the saturator to the condenser. It is expected that this problem can be solved by increasing the thickness of the thermal barrier between the saturator and condenser.

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f08

Figure 8(a) Time series of the number concentrations with the MEMS-based CPC and aerosol electrometer. (b) One-to-one comparison of the measured number concentrations for both systems.

Download

5.4 Detectable concentration range

Even if filtered air is introduced into the system, droplets may form in the condenser via homogeneous or ion-induced nucleation. Droplets without UFP nuclei cause false counting, which makes the system read a higher concentration than what is actually present. This phenomenon is critical, especially in low-concentration environments. To evaluate the false counting of our system, it was operated for 1 h with a HEPA filter connected to its inlet. When the temperature difference between the saturator and condenser was set to 30 C, the average number concentration and counting rate during the measurement period (background concentration) were 0.05 cm−3 and 0.125 s−1, respectively, indicating that homogeneous nucleation hardly occurred. Thus, the temperature profile was uniformly established inside the condensation chip because homogeneous nucleation typically occurs at low temperatures in regions where the local saturation ratio is high. Owing to the low false count performance, our system can be applied to monitoring UFPs in clean rooms of class 100 or 1000 whose background concentrations are on the order of 1 cm−3 (Liao et al., 2018). In these environments, the measurement uncertainty based on Poisson statistics is expected to be 10 % in a sampling time of 40 s for a given flow rate (0.15  L min−1= 2.5 cm3 s−1).

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f09

Figure 9Time series of the number concentrations measured with our system and the reference CPC when the concentration and size were varied.

Download

We characterized the maximum detectable number concentration of our system by comparing the number concentration with that of the AE (i.e., reference number concentration). Monodisperse Ag particles with a size of 25 nm were used as test particles, and their number concentration was increased at intervals of 3 min. Figure 9 shows (a) the time series of number concentrations measured with the MEMS-based CPC and an AE and (b) a one-to-one comparison of the measured number concentrations by both systems. As shown in Fig. 9b, relatively large fluctuations were observed at number concentrations of < 1000 cm−3 because of the electronic noise of the AE. However, in the number concentration range of 1000–5000 cm−3, the overall difference between the concentrations of our system and the AE was only 4.1 %, which proves the high accuracy of our system. When the concentrations exceeded 5000 cm−3, the deviation between the measured and reference number concentrations gradually increased. One of the main reasons was the coincidence error. When multiple particles simultaneously passed through the sensing volume, the miniature OPC could not count them separately. The maximum detectable concentration of our system, which was defined as the number concentration at a difference of 20 %, was 7200 cm−3. Thus, when the concentration exceeded 6852 cm−3, the logarithmic function was fitted to the response curve of our system, as expressed in Fig. 8b. When the calibration based on the fitted curve was applied to our system, the average difference between both systems was within 2.8 %.

5.5 Performance comparison with the reference CPC

The MEMS-based CPC was tested in parallel with a reference CPC. The classification voltage of the DMA was changed to introduce monodisperse Ag particles varying in concentration and size into both systems. The total measurement time was 600 s, and the measured data were averaged in intervals of 6 s. Figure 8 shows the measured number concentrations of our system and the reference CPC. When particles larger than the minimum detectable size (12.9 nm) were introduced, our system clearly showed high accuracy and precision comparable to those of the reference CPC: a difference of 4.54 % at a low concentration (7.99 cm−3 at 28 nm) and a −9.12 % difference at a high concentration (4544.82 cm−3 at 16 nm).

https://www.atmos-meas-tech.net/12/5335/2019/amt-12-5335-2019-f10

Figure 10Time series of the number concentrations measured by our system when it was tilted.

Download

Figure 10 shows the measurement results of our system when it was tilted as shown in the inset. Monodisperse 25 nm Ag particles were introduced, and their concentrations were increased in steps from 0 to 4000 cm−3. Since the measurement was carried out for about 500 s at each angle, the measurement uncertainty of each section was below 0.01 %. When our system was oriented perpendicular to the surface, the difference in counting efficiencies between our system and the reference CPC was 2.04 %, which was similar to the result of the size-dependent counting efficiency. When a 30 angle was applied, the difference in counting efficiency was 7.07 %. At 60, the measurement difference compared to the reference CPC exceeded 10 % (16.3 %). Thus, it was found that, at a tilt angle of 60 or less, the MEMS-based CPC can monitor UFPs without significant degradation in accuracy.

The deviation of the counting efficiency induced from applying a tilt angle can be explained by the sedimentation of droplets in the condenser. At 0, because the direction of gravity was identical to the direction of the sample flow, the probability that grown droplets hit the condenser wall via sedimentation was negligible. However, with an increasing tilt angle, the velocity vector of a droplet perpendicular to the channel increased, which led to a decrease in counting efficiency.

6 Conclusions

The MEMS-based CPC was developed for sensitive and precise monitoring of UFPs at a particular point of interest. Our system comprises two parts: the MEMS-based condensation chip and a miniature OPC. To achieve compactness, the key elements for growing droplets (i.e., the saturator, reservoir, and condenser) were integrated on a 52.5 mm × 60 mm glass slide through a simple photolithographic process and 3-D printing. Quantitative experiments with an AE (model 3068B, TSI Inc., USA) and CPC (model 3772, TSI Inc., USA) demonstrated that our system can count UFPs with a size of 12.9 nm and measure the number concentration with high accuracy (within 4.1 % difference compared to AE) in the range of 1000–5000 cm3.

In terms of compactness and cost-efficiency, our system is superior to conventional instruments. The physical volume of our system is only 8.5 % of the volume of the commercially available portable CPC (e.g., model 3007, TSI Inc., USA). Furthermore, its manufacturing cost can be minimized owing to batch fabrication based on MEMS technology. These advantages allow the system to be successfully applied to various fields that require UFP monitoring. Furthermore, if combined with the recently developed miniature DMA, our system can help realize a mini-scanning mobility particle sizer (mini-SMPS) for an accurate and precise measurement of the UFP size distribution at particular points of interest.

Data availability

Data are available from the authors upon request.

Supplement

The supplement related to this article is available online at: https://doi.org/10.5194/amt-12-5335-2019-supplement.

Author contributions

SJY, HBK, and YJK conceptualized the study. HBK, USH, SJY, DHK, and SML performed the experimental work. SJY wrote the original draft, and SJY, HBK, and DHK revised it. JaH and JuH provided the resources. All authors discussed and commented on the manuscript and approve of its content.

Competing interests

The authors declare that they have no conflict of interest.

Acknowledgements

This work was supported by the Samsung Research Funding & Incubation Center of Samsung Electronics (grant no. SRFC-TA1803-05) and by the Technology Innovation Program (grant no. 10077651, Development of IoT fusion sensor system based on artificial intelligence) funded by the Ministry of Trade, Industry and Energy (MOTIE, Republic of Korea).

Financial support

This work was supported by the Samsung Research Funding & Incubation Center of Samsung Electronics under Project Number SRFC-TA1803-05. This material is based upon work supported by the Ministry of Trade, Industry and Energy (MOTIE, Republic of Korea) under Industrial Technology Innovation Program Number 10077651.

Review statement

This paper was edited by Joachim Curtius and reviewed by three anonymous referees.

References

Chang, H.-K. and Kim, Y.-K.: UV-LIGA process for high aspect ratio structure using stress barrier and C-shaped etch hole, Sens. Actuat. A, 84, 342–350, 2000. 

Chen, X., Wang, X.-P., and Xu, X.: Effective contact angle for rough boundary, Physica D, 242, 54–64, 2013. 

Choi, K.-M., Kim, J.-H., Park, J.-H., Kim, K.-S., and Bae, G.-N.: Exposure characteristics of nanoparticles as process by-products for the semiconductor manufacturing industry, J. Occup. Environ. Hyg., 12, D153–D160, 2015. 

Donaldson, K., Li, X., and MacNee, W.: Ultrafine (nanometre) particle mediated lung injury, J. Aerosol Sci., 29, 553–560, 1998. 

Donovan, R., Locke, B., Osburn, C., and Caviness, A.: Ultrafine aerosol particles in semiconductor cleanrooms, J. Electrochem. Soc., 132, 2730–2738, 1985. 

Hajjam, A., Wilson, J. C., Rahafrooz, A., and Pourkamali, S.: Fabrication and characterization of resonant aerosol particle mass sensors, 2010 IEEE 23rd International Conference on Micro Electro Mechanical Systems (MEMS), Wanchai, Hong Kong, China, 24–28 January 2010, 863–866, https://doi.org/10.1109/MEMSYS.2010.5442343, 2010. 

Hesterberg, T. W., Long, C. M., Bunn, W. B., Lapin, C. A., McClellan, R. O., and Valberg, P. A.: Health effects research and regulation of diesel exhaust: an historical overview focused on lung cancer risk, Inhal. Toxicol., 24 Suppl. 1, 1–45, 2012. 

Hext, P.: Current perspectives on particulate induced pulmonary tumours, Human Experiment. Toxicol., 13, 700–715, 1994. 

Hristozov, D. and Malsch, I.: Hazards and risks of engineered nanoparticles for the environment and human health, Sustainability, 1, 1161–1194, 2009. 

Kang, J. S., Lee, K. S., Lee, K. H., Sung, H. J., and Kim, S. S.: Characterization of a microscale cascade impactor, Aerosol Sci. Technol., 46, 966–972, 2012. 

Kangasluoma, J., Kuang, C., Wimmer, D., Rissanen, M. P., Lehtipalo, K., Ehn, M., Worsnop, D. R., Wang, J., Kulmala, M., and Petäjä, T.: Sub-3 nm particle size and composition dependent response of a nano-CPC battery, Atmos. Meas. Tech., 7, 689–700, https://doi.org/10.5194/amt-7-689-2014, 2014. 

Kangasluoma, J., Hering, S., Picard, D., Lewis, G., Enroth, J., Korhonen, F., Kulmala, M., Sellegri, K., Attoui, M., and Petäjä, T.: Characterization of three new condensation particle counters for sub-3 nm particle detection during the Helsinki CPC workshop: the ADI versatile water CPC, TSI 3777 nano enhancer and boosted TSI 3010, Atmos. Meas. Tech., 10, 2271–2281, https://doi.org/10.5194/amt-10-2271-2017, 2017. 

Kim, H.-L., Han, J. S., Lee, S.-M., Kwon, H.-B., Hwang, J., and Kim, Y.-J.: Ultrafine particle counter using a MEMS-based particle processing chip, 2015 28th IEEE International Conference on Micro Electro Mechanical Systems (MEMS), Estoril, Portugal, 18–22 January 2015, 559–562, https://doi.org/10.1109/MEMSYS.2015.7051016, 2015. 

Kim, H.-L., Han, J., Lee, S.-M., Kwon, H.-B., Hwang, J., and Kim, Y.-J.: MEMS-based particle detection system for measuring airborne ultrafine particles, Sens. Actuat. A, 283, 235–244, 2018. 

Kim, J.-S., Kang, J.-W., and Kim, J.-J.: Simple and low cost fabrication of thermally stable polymeric multimode waveguides using a UV-curable epoxy, Jpn. J. Appl. Phys., 42, 1277, https://doi.org/10.1143/jjap.42.1277, 2003. 

Kim, K. H., Sekiguchi, K., Kudo, S., and Sakamoto, K.: Characteristics of Atmospheric Elemental Carbon (Char and Soot) in Ultrafine and Fine Particles in a Roadside Environment, Japan, Aerosol Air Qual. Res., 11, 1–12, 2011. 

Kittelson, D. B.: Engines and nanoparticles: a review, J. Aerosol Sci., 29, 575–588, 1998. 

Li, N., Sioutas, C., Cho, A., Schmitz, D., Misra, C., Sempf, J., Wang, M., Oberley, T., Froines, J., and Nel, A.: Ultrafine particulate pollutants induce oxidative stress and mitochondrial damage, Environ. Health Perspect., 111, 455–460, 2003. 

Li, N., Georas, S., Alexis, N., Fritz, P., Xia, T., Williams, M. A., Horner, E., and Nel, A.: A work group report on ultrafine particles (American Academy of Allergy, Asthma & Immunology): Why ambient ultrafine and engineered nanoparticles should receive special attention for possible adverse health outcomes in human subjects, J. Allergy Clin. Immunol., 138, 386–396, 2016. 

Liao, B.-X., Tseng, N.-C., Li, Z., Liu, Y., Chen, J.-K., and Tsai, C.-J.: Exposure assessment of process by-product nanoparticles released during the preventive maintenance of semiconductor fabrication facilities, J. Nanopart. Res., 20, 203, https://doi.org/10.1007/s11051-018-4302-7, 2018. 

Libman, S., Wilcox, D., and Zerfas, B.: Ultrapure Water for Advance Semiconductor Manufacturing: Challenges and Opportunities, ECS Transact., 69, 17–28, 2015. 

Liu, L., Urch, B., Poon, R., Szyszkowicz, M., Speck, M., Gold, D. R., Wheeler, A. J., Scott, J. A., Brook, J. R., Thorne, P. S., and Silverman, F. S.: Effects of ambient coarse, fine, and ultrafine particles and their biological constituents on systemic biomarkers: a controlled human exposure study, Environ. Health Perspect., 123, 534–540, 2015. 

Liu, Q. and Chen, D.-R.: Experimental evaluation of miniature plate DMAs (mini-plate DMAs) for future ultrafine particle (UFP) sensor network, Aerosol Sci. Technol., 50, 297–307, 2016. 

Manodori, L. and Benedetti, A.: Nanoparticles monitoring in workplaces devoted to nanotechnologies, Nanosafe 2008: International Conference on Safe production and use of nanomaterials, 170, 012001, https://doi.org/10.1109/MEMSYS.2010.5442343,2009. 

McMurry, P. H.: The history of condensation nucleus counters, Aerosol Sci. Technol., 33, 297–322, 2000. 

Neisser, M. and Wurm, S.: ITRS lithography roadmap: 2015 challenges, Adv. Opt. Technol., 4, 235–240, 2015.  

Qi, C. and Kulkarni, P.: Miniature differential mobility analyzer for compact field-portable spectrometers, Aerosol Sci. Technol., 50, 1145–1154, 2016. 

Renwick, L., Brown, D., Clouter, A., and Donaldson, K.: Increased inflammation and altered macrophage chemotactic responses caused by two ultrafine particle types, Occup. Environ. Med., 61, 442–447, 2004. 

Shi, J. P., Khan, A., and Harrison, R. M.: Measurements of ultrafine particle concentration and size distribution in the urban atmosphere, Sci. Total Environ., 235, 51–64, 1999. 

Sioutas, C.: Evaluation of the measurement performance of the scanning mobility particle sizer and aerodynamic particle sizer, Aerosol Sci. Technol., 30, 84–92, 1999. 

Sousan, S., Koehler, K., Hallett, L., and Peters, T. M.: Evaluation of the Alphasense Optical Particle Counter (OPC-N2) and the Grimm Portable Aerosol Spectrometer (PAS-1.108), Aerosol Sci. Technol., 50, 1352–1365, 2016. 

Stansbury, J. W. and Idacavage, M. J.: 3-D printing with polymers: Challenges among expanding options and opportunities, Dental Mat., 32, 54–64, 2016. 

Stolzenburg, D., Steiner, G., and Winkler, P. M.: A DMA-train for precision measurement of sub-10  nm aerosol dynamics, Atmos. Meas. Tech., 10, 1639–1651, https://doi.org/10.5194/amt-10-1639-2017, 2017. 

Stolzenburg, M. R. and McMurry, P. H.: An ultrafine aerosol condensation nucleus counter, Aerosol Sci. Technol., 14, 48–65, 1991. 

Wasisto, H. S., Merzsch, S., Waag, A., Uhde, E., Salthammer, T., and Peiner, E.: Airborne engineered nanoparticle mass sensor based on a silicon resonant cantilever, Sens. Actuat. B, 180, 77–89, 2013. 

Zhang, C., Zhu, R., and Yang, W.: A micro aerosol sensor for the measurement of airborne ultrafine particles, Sensors, 16, 399, https://doi.org/10.3390/s16030399, 2016. 

Download
Short summary
We present a portable, inexpensive, and accurate microelectromechanical-system-based (MEMS-based) condensation particle counter (CPC) for the sensitive and precise monitoring of airborne ultrafine particles (UFPs). The CPC is miniaturized by utilizing MEMS technology and 3-D printing. Thus, the proposed system can potentially be used for UFP monitoring in various environments.