Journal cover Journal topic
Atmospheric Measurement Techniques An interactive open-access journal of the European Geosciences Union
Journal topic

Journal metrics

Journal metrics

  • IF value: 3.400 IF 3.400
  • IF 5-year value: 3.841 IF 5-year
    3.841
  • CiteScore value: 3.71 CiteScore
    3.71
  • SNIP value: 1.472 SNIP 1.472
  • IPP value: 3.57 IPP 3.57
  • SJR value: 1.770 SJR 1.770
  • Scimago H <br class='hide-on-tablet hide-on-mobile'>index value: 70 Scimago H
    index 70
  • h5-index value: 49 h5-index 49
Volume 9, issue 2
Atmos. Meas. Tech., 9, 553–576, 2016
https://doi.org/10.5194/amt-9-553-2016
© Author(s) 2016. This work is distributed under
the Creative Commons Attribution 3.0 License.
Atmos. Meas. Tech., 9, 553–576, 2016
https://doi.org/10.5194/amt-9-553-2016
© Author(s) 2016. This work is distributed under
the Creative Commons Attribution 3.0 License.

Research article 17 Feb 2016

Research article | 17 Feb 2016

A two-channel thermal dissociation cavity ring-down spectrometer for the detection of ambient NO2, RO2NO2 and RONO2

J. Thieser1, G. Schuster1, J. Schuladen1, G. J. Phillips1,a, A. Reiffs1, U. Parchatka1, D. Pöhler2, J. Lelieveld1, and J. N. Crowley1 J. Thieser et al.
  • 1Max-Planck-Institut für Chemie, Division of Atmospheric Chemistry, Mainz, Germany
  • 2Institute of Environmental Physics, University of Heidelberg, Heidelberg, Germany
  • anow at: Department of Natural Sciences, University of Chester, Chester, UK

Abstract. We describe a thermal dissociation cavity ring-down spectrometer (TD-CRDS) for measurement of ambient NO2, total peroxy nitrates (ΣPNs) and total alkyl nitrates (ΣANs). The spectrometer has two separate cavities operating at  ∼  405.2 and 408.5 nm. One cavity (reference) samples NO2 continuously from an inlet at ambient temperature, the other samples sequentially from an inlet at 473 K in which PNs are converted to NO2 or from an inlet at 723 K in which both PNs and ANs are converted to NO2, difference signals being used to derive mixing ratios of ΣPNs and ΣANs. We describe an extensive set of laboratory experiments and numerical simulations to characterise the fate of organic radicals in the hot inlets and cavity and derive correction factors to account for the bias resulting from the interaction of peroxy radicals with ambient NO and NO2. Finally, we present the first measurements and comparison with other instruments during a field campaign, outline the limitations of the present instrument and provide an outlook for future improvements.

Publications Copernicus
Download
Short summary
We report on the use of thermal dissociation cavity ring-down spectroscopy to detect NO2, peroxy nitrates and alkyl nitrates. We present both laboratory studies that characterise the chemical formation and loss of NO2 in the heated inlets and also result from a first field deployment.
We report on the use of thermal dissociation cavity ring-down spectroscopy to detect NO2, peroxy...
Citation